Software
​
Current version: 6.0.1
Software release notes
New features
High MP values – this release includes an update to deliver a better user experience for patients with very high MPOD scores. When the patient’s MPOD value was high AND no point of isoluminance could be determined before the test ended, the device previously delivered a Reject message. In Version 5.2, if the patient has not reached isoluminance by the end of the test, the software will now deliver a Test Successful message and provide the MPOD value as > 0.9, providing the data is not noisy and conforms to the Confidence Level algorithm.
Importing databases from v2.03H – some customers have been experiencing difficulty importing old patient databases that predate the Confidence Level feature enhancements that were available in Version 4 onward.
Please note that patients’ existing records will not have confidence values calculated if this was not in the original test record.
Databases created using the MPS II v4.xx application are imported into the new Version 5 format when the application is run for the first time after installation. Databases created using the MPS II v2.03H application are imported into the new Version 5 format when the application is run for the first time after installation.
Additionally, in this release the importing of legacy databases – usually version 2.03H and earlier – is now included in the installer routine. This removes the previous need to install in two steps.
​
Bug fixes
Minor modification to data smoothing algorithm and display.
Supported platforms
Windows 7 Professional, and above.
Software dependencies
Adobe Reader XI must be installed for viewing reports.

Navigate More about MPS
​